Vaisala Humidity and Temperature Probe HMP155

Features

• Vaisala HUMICAPâ180R sensor: superior long-term stability

• Optional warmed humidity probe and chemical purge

• Plug-and-play

• USB connection for service use

• Use with DTR13 and DTR503 radiation shields and a Stevenson screen

• Weather-proof housing IP66

• Optional, fast temperature probe

• Different output possibilities: voltage, RS-485, resistive Pt100

• Applications: meteorology, aviation and road weather, instrumentation

 

Click here for datasheet:

HMP155-Datasheet-B210752EN-F

Click here for reference guide:

HMP155 Quick Reference Guide – M210913EN-C

The Vaisala Humicap HMP155 humidity and temperature probe provides reliable humidity and temperature measurement. The HMP155 has a new generation Vaisala Humicap®180R sensor that has excellent stability and withstands harsh environments. The probe structure is solid and the sensor is protected with a sintered teflon filter, which gives maximum protection against liquid water, dust, and dirt.

Measuring humidity reliably is challenging in environments where humidity is near saturation. Measurements may be corrupted by fog, mist, rain, and heavy dew. A wet probe may not give an accurate measurement in the ambient air.

This is an environment to which Vaisala has designed a patented, warmed probe for reliable measuring. As the sensor head is warmed continuously, the humidity level inside it stays below the ambient level. Thus, it also reduces the risk of condensation forming on the probe.

The HMP155 is especially designed for use in meteorological applications, such as synoptic and hydrological weather stations, aviation, and road weather.

Technical Specifications

RELATIVE HUMIDITY

Measurement range 0 – 100 %RH
Accuracy (incl. non-linearity, hysteresis and repeatability) at
+15 to +25 °C (+59 to +77 °F)
-20 to+40 °C (-4 to 104 °F)
-40 to -20 °C (-40 to -4 °F)
+40 to +60 °C (+104 to +140 °F)
-60 to -40 °C (-76 to -40 °F)
Factory calibration
uncertainty (+20 °C /+68 °F)
±1 %RH (0 – 90 %RH)
±1.7 %RH (90 – 100 %RH)
±(1.0 + 0.008 x reading) %RH
±(1.2 + 0.012 x reading) %RH
±(1.2 + 0.012 x reading) %RH
±(1.4 + 0.032 x reading) %RH
±0.6 %RH (0 to 40 %RH)*
±1.0 %RH (40 … 97 %RH)*
Response time at +20 °C in still air with a sintered PTFE filter
63%
90%
20 s
60 s

TEMPERATURE

Measurement range -80 °C to +60 °C (-112 °F to +140 °F)
Accuracy with voltage output at
-80 to +20 °C
+20 to +60 °C
passive (resistive) output
according to IEC 751 1/3 Class B
RS485 output
-80 to +20 °C
+20 to +60 °C
±(0.226 – 0.0028 x temperature) °C
±(0.055 + 0.0057 x temperature) °C
±(0.1 + 0.00167 x |temperature|)°C
±(0.176 – 0.0028 x temperature) °C
±(0.07 + 0.0025 x temperature) °C
Temperature sensor Pt100 RTD Class F0.1 IEC 60751
Response time with additional temperature probe in 3 m/s air flow
63%
90%
20 s
35 s

GENERAL

Operating temperature range -80 °C to +60 °C (-112 °F to +140 °F)
Storage temperature range -80 °C to +60 °C (-112 °F to +140 °F)
Connection 8-pin male M12 connector
Connection cables 3.5, 10, and 30 m
Cable material PUR
Wire size AWG26
Service cables USB connection cable
MI70 connection cable
Additional T probe cable length 2 m
Housing material PC
Housing classification IP66
Sensor protection sintered PTFE
optional membrane filter
Weight (probe) 86 g
Dimensions 267 mm L x 40 mm Ø
Electromagnetic compatibility Complies with the EMC standard EN61326-1, Electrical equipment for measurement control and laboratory use – EMC requirement for use in industrial locations

INPUTS AND OUTPUTS

Operating voltage 7 to 28 VDC
Note: minimum operating voltage 12 V with 0 to 5 V output and 16 V with 0 to 10 V output, probe heating, chemical purge or XHEAT.
Outputs voltage output (0 – 1 V, 0 – 5 V, 0 – 10 V)
resistive Pt100 (4-wire connection)
RS485
Average current consumption (+15 VDC, load 100 kOhm)

0 to 1 V output

0 to 10 V output

RS485

during chemical purge

with warmed probe

3 mA
+0.5 mA
4 mA
max 110 mA
max. 150 mA
Settling time at power-up
voltage output
RS485
2 s
3s